Specifications - 2100 Bioanalyzer RNA Kits

  RNA 6000 Nano total RNA Kit RNA 6000 Nano mRNA Kit RNA 6000 Pico Total RNA Kit RNA 6000 Pico mRNA Kit Small RNA Kit
Analytical Specifications
Quantitative range 25–500 ng/µL 25–250 ng/µL - - 50-2000 pg/µL of purified miRNA in water
Qualitative range 5–500 ng/µL 25–250 ng/µL 50–5000 pg/µL in water 250–5000 pg/µL in water 50-2000 pg/µL of purified miRNA in water
Sensitivity (S/N>3) 5 ng/µL in water 25 ng/µL in water 50 pg/µL in water or 200 pg/µL in TE 250 pg/µL in water or 500 pg/µL in TE 50 pg/µL in water**
Quantitation reproducibility 10% CV (within a chip) 10% CV (within a chip) 20% CV (within a chip) 20% CV (within a chip) 25% CV (within a chip)
Quantitation accuracy 20% CV* 20% CV* 30% CV* 30% CV* -
Maximum sample buffer strength 100 mM Tris
0.1 mM EDTA
or 125 mM NaCl
or 15 mM MgCl2
100 mM Tris
0.1 mM EDTA
or 125 mM NaCl
or 15 mM MgCl2
50 mM Tris
0.1 mM EDTA
or 50 mM NaCl
or 15 mM MgCl2
50 mM Tris
0.1 mM EDTA
or 50 mM NaCl
or 15 mM MgCl2
10 mM Tris
0.1 mM EDTA



  RNA 6000 Nano total RNA Kit RNA 6000 Nano mRNA Kit RNA 6000 Pico total RNA Kit RNA 6000 Pico mRNA Kit Small RNA Kit
Physical Specifications
Analysis time 30 minutes 30 minutes 30 minutes 30 minutes 30 minutes
Samples per chip 12 12 11 11 11
Sample volume 1 µL 1 µL 1 µL 1 µL 1 µL
Kit stability ≥4 months at 4°C ≥4 months at 4°C ≥4 months at 4°C ≥4 months at 4°C ≥4 months at 4°C
25 chips per kit RNA Nano 12/chip = 300 samples/kit RNA Pico 11/chip = 275 samples/kit -



  * Determined analyzing the RNA ladder as sample
** Measeured for the 40 nt fragment of the Small RNA

For Research Use Only. Not for use in diagnostic procedures